User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP 44720:Strain Multiplexer DATE: 19-Sep-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 12 NUMBER OF LINES: 220 CONFIGURATION: Fluke 5700A CONFIGURATION: HP 34401A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 HEAD INITIAL CONDITIONS 1.004 DISP The following Standard test equipment is required: 1.004 DISP Measurements Group 1550A Strain Indicator Calibrator. 1.005 HEAD {* Strain Measurements *} 1.006 JMP 2.001 1.007 EVAL 2.001 DISP Turn off UUT. Remove UUT adaptor block, if necessary. 2.002 DISP If necessary place the UUT card to tested into 2.002 DISP mainfarme SLOT 3. 2.003 DISP Connect the 44720A Test Adaptor. 2.004 DISP Connect the UUT to the 1550A as follows: 2.004 DISP UUT 1550A 2.004 DISP P+ ............................. 350 ohm P+ 2.004 DISP P- ............................. 350 ohm P- 2.004 DISP S+ ............................. 350 ohm S+ 2.004 DISP S- ............................. 350 ohm S- 2.005 DISP Use guards and grounds as neccesary. 2.006 DISP Connect the 34401A to the 1550A as follows: 2.006 DISP 34401A 1550A 2.006 DISP INPUT HI ....................... 350 ohm P+ 2.006 DISP INPUT LO ....................... 350 ohm P1 2.007 DISP Connect the UUT to the 5700A as follows: 2.007 DISP UUT 5700A 2.007 DISP VOLTS IN + ..................... OUTPUT HI V ohm A 2.007 DISP VOLTS IN - ..................... OUTPUT LO V ohm A 2.008 DISP Depress the - POLARITY on the 1550A. 2.009 DISP Apply power to the 3852A Mainframe and wait for the 2.009 DISP 3852A to complete it's power-up self test. 2.010 IEEE [@34401]FUNC "VOLT:DC"[GTL] 2.011 DISP Set the STRAIN CALIBRATOR to 0,0,0. 2.012 MESS Adjust the 5700 for a reading of 5.000V on the 34401A. 2.013 5700 5V N 2W 2.014 MEME >1 2.015 MESS 2.016 MEME <1 2.017 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 2.018 IEEE [SDC] 2.019 IEEE [TERM LF] 2.020 IEEE RST; USE 600; REAL FB_REF 2.021 IEEE CONFMEAS STRUN,300,INTO FB_REF 2.022 DISP Set the 1550A 9,9,9. 2.023 IEEE CONFMEAS STRQ,300,REF FB_REF 2.024 IEEE [I] 2.025 MEM* 1000000 2.026 ACC 99900.0µE 0.025% 2.027 MEME 2.028 MEMC µE 20.0% ch0 #! WARNING: Test Tol 49.95, Sys Tol 12.488, TUR 4.000 (< 4.00). 3.001 DISP Set the STRAIN CALIBRATOR to 0,0,0. 3.002 MEME <1 3.003 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 3.004 IEEE [SDC] 3.005 IEEE [TERM LF] 3.006 IEEE RST; USE 600; REAL FB_REF 3.007 IEEE CONFMEAS STRUN,301,INTO FB_REF 3.008 DISP Set the 1550A 9,9,9. 3.009 IEEE CONFMEAS STRQ,301,REF FB_REF 3.010 IEEE [I] 3.011 MEM* 1000000 3.012 ACC 99900.0µE 0.025% 3.013 MEME 3.014 MEMC µE 20.0% ch1 #! WARNING: Test Tol 49.95, Sys Tol 12.488, TUR 4.000 (< 4.00). 4.001 DISP Set the STRAIN CALIBRATOR to 0,0,0. 4.002 MEME <1 4.003 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 4.004 IEEE [SDC] 4.005 IEEE [TERM LF] 4.006 IEEE RST; USE 600; REAL FB_REF 4.007 IEEE CONFMEAS STRUN,302,INTO FB_REF 4.008 DISP Set the 1550A 9,9,9. 4.009 IEEE CONFMEAS STRQ,302,REF FB_REF 4.010 IEEE [I] 4.011 MEM* 1000000 4.012 ACC 99900.0µE 0.025% 4.013 MEME 4.014 MEMC µE 20.0% ch2 #! WARNING: Test Tol 49.95, Sys Tol 12.488, TUR 4.000 (< 4.00). 5.001 DISP Set the STRAIN CALIBRATOR to 0,0,0. 5.002 MEME <1 5.003 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 5.004 IEEE [SDC] 5.005 IEEE [TERM LF] 5.006 IEEE RST; USE 600; REAL FB_REF 5.007 IEEE CONFMEAS STRUN,303,INTO FB_REF 5.008 DISP Set the 1550A 9,9,9. 5.009 IEEE CONFMEAS STRQ,303,REF FB_REF 5.010 IEEE [I] 5.011 MEM* 1000000 5.012 ACC 99900.0µE 0.025% 5.013 MEME 5.014 MEMC µE 20.0% ch3 #! WARNING: Test Tol 49.95, Sys Tol 12.488, TUR 4.000 (< 4.00). 6.001 DISP Set the STRAIN CALIBRATOR to 0,0,0. 6.002 MEME <1 6.003 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 6.004 IEEE [SDC] 6.005 IEEE [TERM LF] 6.006 IEEE RST; USE 600; REAL FB_REF 6.007 IEEE CONFMEAS STRUN,304,INTO FB_REF 6.008 DISP Set the 1550A 9,9,9. 6.009 IEEE CONFMEAS STRQ,304,REF FB_REF 6.010 IEEE [I] 6.011 MEM* 1000000 6.012 ACC 99900.0µE 0.025% 6.013 MEME 6.014 MEMC µE 20.0% ch4 #! WARNING: Test Tol 49.95, Sys Tol 12.488, TUR 4.000 (< 4.00). 7.001 DISP Set the STRAIN CALIBRATOR to 0,0,0. 7.002 MEME <1 7.003 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 7.004 IEEE [SDC] 7.005 IEEE [TERM LF] 7.006 IEEE RST; USE 600; REAL FB_REF 7.007 IEEE CONFMEAS STRUN,305,INTO FB_REF 7.008 DISP Set the 1550A 9,9,9. 7.009 IEEE CONFMEAS STRQ,305,REF FB_REF 7.010 IEEE [I] 7.011 MEM* 1000000 7.012 ACC 99900.0µE 0.025% 7.013 MEME 7.014 MEMC µE 20.0% ch5 #! WARNING: Test Tol 49.95, Sys Tol 12.488, TUR 4.000 (< 4.00). 8.001 DISP Set the STRAIN CALIBRATOR to 0,0,0. 8.002 MEME <1 8.003 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 8.004 DISP Set the STRAIN CALIBRATOR to 0,0,0. 8.005 MEME <1 8.006 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 8.007 IEEE [SDC] 8.008 IEEE [TERM LF] 8.009 IEEE RST; USE 600; REAL FB_REF 8.010 IEEE CONFMEAS STRUN,306,INTO FB_REF 8.011 DISP Set the 1550A 9,9,9. 8.012 IEEE CONFMEAS STRQ,306,REF FB_REF 8.013 IEEE [I] 8.014 MEM* 1000000 8.015 ACC 99900.0µE 0.025% 8.016 MEME 8.017 MEMC µE 20.0% ch6 #! WARNING: Test Tol 49.95, Sys Tol 12.488, TUR 4.000 (< 4.00). 9.001 DISP Set the STRAIN CALIBRATOR to 0,0,0. 9.002 MEME <1 9.003 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 9.004 IEEE [SDC] 9.005 IEEE [TERM LF] 9.006 IEEE RST; USE 600; REAL FB_REF 9.007 IEEE CONFMEAS STRUN,307,INTO FB_REF 9.008 DISP Set the 1550A 9,9,9. 9.009 IEEE CONFMEAS STRQ,307,REF FB_REF 9.010 IEEE [I] 9.011 MEM* 1000000 9.012 ACC 99900.0µE 0.025% 9.013 MEME 9.014 MEMC µE 20.0% ch7 #! WARNING: Test Tol 49.95, Sys Tol 12.488, TUR 4.000 (< 4.00). 10.001 DISP Set the STRAIN CALIBRATOR to 0,0,0. 10.002 MEME <1 10.003 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 10.004 IEEE [SDC] 10.005 IEEE [TERM LF] 10.006 IEEE RST; USE 600; REAL FB_REF 10.007 IEEE CONFMEAS STRUN,308,INTO FB_REF 10.008 DISP Set the 1550A 9,9,9. 10.009 IEEE CONFMEAS STRQ,308,REF FB_REF 10.010 IEEE [I] 10.011 MEM* 1000000 10.012 ACC 99900.0µE 0.025% 10.013 MEME 10.014 MEMC µE 20.0% ch8 #! WARNING: Test Tol 49.95, Sys Tol 12.488, TUR 4.000 (< 4.00). 11.001 DISP Set the STRAIN CALIBRATOR to 0,0,0. 11.002 MEME <1 11.003 5700 V S 2W #! System Uncertainty not calculated because NOMINAL & RANGE values not spec. 11.004 IEEE [SDC] 11.005 IEEE [TERM LF] 11.006 IEEE RST; USE 600; REAL FB_REF 11.007 IEEE CONFMEAS STRUN,309,INTO FB_REF 11.008 DISP Set the 1550A 9,9,9. 11.009 IEEE CONFMEAS STRQ,309,REF FB_REF 11.010 IEEE [I] 11.011 MEM* 1000000 11.012 ACC 99900.0µE 0.025% 11.013 MEME 11.014 MEMC µE 20.0% ch9 12.001 IEEE [SDC] 12.002 IEEE [@34401]*RST #! WARNING: Test Tol 49.95, Sys Tol 12.488, TUR 4.000 (< 4.00). 12.003 END #! T.U.R.s less than 4.00: 11 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.